Technical data


Qualitative X-ray fluorescence analysis of powder samples

Article no: P2544705

Principle

Various powder samples are subjected to polychromatic X-rays. The energy of the resulting fluorescence radiation is analysed with the aid of a semiconductor detector and a multichannel analyser. The energy of the corresponding characteristic X-ray fluorescence lines is determined. The elements of the samples are identified by comparing the line energies with the corresponding table values.

Tasks

  1. Calibrate the semiconductor energy detector with the aid of the characteristic radiation of the tungsten X-ray tube.
  2. Record the fluorescence spectra that are produced by the samples.
  3. Determine the energy values of the corresponding fluorescence lines and compare the experimental energy values with the corresponding table values in order to identify the powder components.

Learning objectives

  • Bremsstrahlung
  • Characteristic X-radiation
  • Energy levels
  • Fluorescent yield
  • Semiconductor energy detectors
  • Multichannel analyse

Software included. Computer not provided.

Scope of delivery

XR 4.0 expert unit, 35 kV 09057-99 1
XR 4.0 X-ray goniometer 09057-10 1
XR4 X-ray Plug-in Cu tube 09057-51 1
XR 4.0 X-ray material upgrade set 09165-88 1
XR 4.0 X-ray Chemical set for edge absorption 09056-07 1