setTimeout(function(){ window.print(); },500)
 

Technical data


Energy loss of alpha particles in gases with MCA

Article no: P2522415

Principle

A study is made of the connection between the energy E of alpha-particles and the path x travelled by them in air at standard pressure. The measurements recorded enable the differencial energy loss dE/dx to be calculated as a function of x.

Tasks

  1. The spectrum of a covered 241Am source is measured at a fixed distance s as a function of the pressure p. The distance s is selected in such a way as to correspond to the maximum range at the highest pressure measurable with the manometer used. The energy corresponding to the central points of the individual spectra are determined (after calibration of the measurement layout with an open 241Am-emitter, see 3.) and plotted as a function of the distance x converted to a 1013 hPa basis. Using this function, the differential energy loss (dE/dx) is then calculated as a function of x and again plotted on the graph.
  2. The spectrum of the source used in 1 is measured initially under the same geometric conditions under vacuum and subsequently with the vessel filled with helium, nitrogen or carbon dioxide, in each case under identical pressures. The different energy loss values are compared with the electron concentration in the particular gas.
  3. The mean energy with which the alpha-particles leave the covered americium source is determined by calibration against the open americium emitter (E = 5.485 MeV). (This value is required for the evaluation in 1.)

Learning objectives

  • Range
  • Range dispersion
  • Mean free path length
  • Mean ionization energy of gas atoms
  • Mean energy loss of a-particles per collision
  • Differencial energy loss
  • Bethe formula
  • Electron concentration in gases

Scope of delivery

PHYWE Multichannel Analyser (MCA) 13727-99 1
measure Software multi channel analyser 14452-61 1
Alpha and Photodetector 09099-00 1
Radioactive source Am-241, 3.7 kBq 09090-03 1
Radioactive source Am-241, 370 kBq 09090-11 1
Container for nuclear physics experiments 09103-00 1
Pre-amplifier for alpha detector 09100-10 1
Vacuum gauge DVR 2 pro, 1 ... 1000 hPa 34171-00 1
Diaphragm pump, two stage, 220V 08163-93 1
Vacuum tube, NBR, 6/14mm, 1 m 39289-00 3
Hose clamp for 10-16 mm diameter 41001-00 9
Tubing connector,Y-shape, ID 8-9mm 47518-03 2
Pinchcock, width 20 mm 43631-20 1
Fine control valve 33499-00 1
Compressed gas, helium, 12 l 41772-03 1
Compressed gas, propane, 7 l 41772-10 1
Compressed gas, CO2, 22 g 41772-06 1
Screened cable, BNC, l = 1 m 07542-11 3

PHYWE Systeme GmbH & Co. KG
Robert-Bosch-Breite 10 – 37079 Göttingen – Germany
www.phywe.com